Read Hiden Analytical MASsoft QMS experiment files directly and export to csv.
MASsoft stores an experiment as a group of files sharing one stem: .scn
(scan definitions), .exp (an index), .dat (the data), .env (the
instrument and its settings), plus .ann. This package decodes them and
hands back the readings, with a massoft2csv command for the common case.
By default the CSV it writes is byte-for-byte what MASsoft itself would
export — same header, same fixed-width columns, same CRLF endings.
This is tested by comparison of exports of the same data by MASsoft
and this package. An example experiment is in tests/data.
uv syncOr add it to another project:
uv add git+https://github.com/blongworth/massoft-exportThe package itself has no dependencies. The polars and pandas extras only
matter if you want to_polars() / to_pandas().
massoft2csv path/to/run.expWrites run.csv beside the experiment. Pass any member file or the bare stem;
pass several at once; use -o for a specific destination or -o - for stdout.
massoft2csv data/*.expA MID experiment — every scan parked on one mass — comes out in MASsoft's own layout: a 24-line header describing the instrument, the scans and the environment, then one fixed-width row per cycle.
0000000437,scans,0000000092,DataLength
"header",0000000024,"lines"
"Date",8/14/2026,"Time",21:20:59
"ID","HAL RC HALO 201 #15876","Version",3
...
"Time","ms","14 amu","32 amu","40 amu","18 amu","44 amu",
00:00:01, 1521, 5.55624e-09, 1.46572e-08, 5.89014e-10, ...
--style tidy drops the header and gives every reading an absolute
timestamp instead of time-since-start:
massoft2csv --style tidy path/to/run.expdatetime,elapsed_ms,14 amu,32 amu,40 amu,18 amu,44 amu
2026-08-14 21:20:59,1521,5.55624e-09,1.46572e-08,5.89014e-10,2.27233e-10,3.91716e-11
An experiment containing a mass sweep is always written this way — long, one row per measured point, with the mass axis reconstructed from the scan definition — because MASsoft's wide layout has no column for a mass axis:
cycle,datetime,elapsed_ms,scan,detector,mass,value
1,2026-08-12 21:30:34,391,1,Faraday,2,2.14999e-10
import massoft_export
run = massoft_export.read_experiment("data/20260814hrs2609f2")
run.columns # ['14 amu', '32 amu', '40 amu', '18 amu', '44 amu']
len(run) # 437 cycles
run.start_time # datetime.datetime(2026, 8, 14, 21, 20, 59)
run.is_mid # True
for time, elapsed_ms, *values in run.rows():
...
df = run.to_polars()
run.write_csv("run.csv") # MASsoft's format
run.write_csv("run.csv", "tidy") # or the plain tablerun.environment holds the instrument identity and its setting modes, which
is where the export header's "ID" line and "Environment" block come from:
run.environment.instrument # 'HAL RC HALO 201 #15876'
run.environment.active_mode.name # 'RGA'
run.environment.active_mode["emission"].value # 50.0run.scans holds the Scan definitions (start, stop, increment, dwell,
settle, detector). run.cycles holds one Cycle per pass, each with a
Reading per scan carrying its own timestamp — the masses in a cycle are
measured seconds apart, and each is stamped when it was taken. For sweeps, or
for anything you would rather have long than wide, run.points() yields every
measured point as a dict.
MASsoft's export gives only time since the start of the run, and the
massoft style reproduces that. The absolute time of every reading is
reconstructed too, and the tidy style writes it — useful for merging
against underway or CTD data. Two things to know:
- Elapsed time and wall-clock time drift apart. The millisecond counter
stops while acquisition is paused; the clock does not. One 17-hour run in
the test set accumulates 153 seconds of drift across four pause/resume
events. Both are reported:
elapsed_msanddatetime. - The instrument clock can be reset mid-run, which on a ship means timestamps step backwards when the vessel changes time zone. That is left as recorded rather than smoothed over. Only a near-full-day step backwards is corrected, since that one is a MASsoft artifact: the seconds counter and the Julian day number roll over midnight one reading apart.
massoft_export/format.py documents the .scn/.exp/.dat layout — block
structure, field offsets, and the quirks worth knowing — and
massoft_export/environment.py documents the .env. Both were worked out by
inspection and checked against MASsoft exports; neither is from a vendor
specification, so treat anything they call "in practice" as an observation
rather than a guarantee.
Two details of the export format are inferred rather than confirmed, because
the available exports do not distinguish the alternatives. Increment is
written with %g, which reproduces every observed value without losing
precision on a finer step. And MASsoft writes the first character of the
scan count as a on some exports and 0 on others, with nothing else in
the file explaining which; this always writes 0, so the field parses as a
number.
uv run pytesttests/data holds two small real experiments — a five-mass MID run and a
2–50 amu Faraday background sweep — with MASsoft's own export of the first as
the regression baseline.